1. Noțiuni de bază despre echipament

Specificații tehnice 1.1

LISUN Model GNGPL-3610 GNGPL-3610-2PA
Numărul de posturi de lucru 1 2
Lamă 1 round blade (Ф6) One rectangular blade and one Ф6 circular blade
Grosimea lamei 0.7 mm
greutăţi 2.5N weight (including the weight of the blade): 1 2.5N weights (including the weight of the blade): 2
Standarde aplicabile BS 1363-1 Figura 10 GB/T 2099.1, Clauza 30.1 și Figura 41 (IEC 60884-1, Clause 30.1 and Figure 44)

Note: This device must be used in conjunction with the LISUN GW-225 High Temperature Chamber and is specifically designed for high-temperature pressure testing of insulation sheaths on plugs and prongs for household and similar purposes.

1.2 Scenarii de aplicare

Plug and Socket Manufacturers: Inspection of incoming plastic materials, 100% final inspection of finished products, and comparative testing of new materials; Third-Party Safety Testing Laboratories: CCC certification, CE (IEC) certification, and mandatory testing for UK BS certification; Quality Inspection / Market Regulatory Authorities: Random inspections of plug products in e-commerce and brick-and-mortar markets; Small Appliance Manufacturers: In-house reliability verification of plugs for adapters, rice cookers, electric kettles, and desk lamps.

1.3 Download Electronic Documents

V-am trimis deja linkurile de descărcare pentru manualul de utilizare electronic, certificatul de calibrare, cardul de garanție și alte documente prin e-mail în momentul expedierii. Vă rugăm să le descărcați cât mai curând posibil. Dacă nu ați primit e-mailul, vă rugăm să ne contactați pentru a solicita din nou linkurile de descărcare.

2. Măsuri de siguranță

During high-temperature testing, temperatures may reach up to 200°C. When opening the door to remove or place test specimens, wear heat-resistant gloves; do not touch the inner chamber, blades, weights, or fixtures with bare hands to prevent burns.

Weights are heavy metal objects; handle them gently. Do not throw or stack them. The blades are sharp; be careful of the cutting edges when installing or replacing test specimens to avoid cutting your hands. When moving the entire press fixture, lift it steadily to prevent the blades from causing scratches.

3. Prezentare generală a echipamentului

3.1 Diagramă standard

Vezi Figura 3-1 (preluată din IEC 60884-1, Figura 44).

High Temperature Pressure Test Device-Figure1

Figura 3-1

3.2 Photographs of the Actual Items

3.2.1 GNGPL-3610

Vezi Figura 3-2.

Număr notițe
Lamă rotundă
greutăţi
Baza reglabilă

High Temperature Pressure Test Device-Figure2

Figura 3-2

3.2.2 GNGPL-3610-2PA

Vezi Figura 3-3.

Număr notițe
Rectangular Blade
Lamă rotundă
③④ greutăţi
Baza reglabilă

High Temperature Pressure Test Device-Figure3

Figura 3-3

4. High-Temperature Pressure Test of the Insulation Sheath

4.1 Preparation of Specimens and Apparatus

Rectangular blades are used for round-pin tests; circular blades are used for other pin tests. Refer to Section 3.1 to position the specimen correctly.

4.2 Procedura de testare

Set the high-temperature chamber to 200°C and allow it to stabilize at the target temperature. Place the test setup, along with the test specimen mounted in its normal position, in the high-temperature chamber for 2 hours.

Next, remove the test apparatus and the specimen, detach the specimen from the apparatus, and immerse it in cold water within 10 seconds to cool it.

4.3 Evaluarea rezultatelor testelor

Measure the thickness of the insulating material at the indentation point; the reduction in thickness compared to the original measurement at the start of the test should not exceed 50 percent.

4.4 Proceduri post-testare

4.4.1 Resetare dispozitiv

After the test is complete, remove the weights and blades from the test rig bracket. Do not allow the weights and blades to remain suspended on the test rig for extended periods when the test is not in progress.

4.4.2 Pregătirea probei

Separați probele care au trecut testul de cele care nu au trecut de acesta, etichetați-le clar și evitați orice confuzie.

5. Întreținerea și îngrijirea zilnică a echipamentelor

5.1 Întreținerea zilnică după oprire

At the end of each day’s testing, clean dust and sample debris from the equipment surfaces, and keep the chucks and slide rails clean and free of debris; check that the blades are not deformed.

5.2 Întreținerea în timpul depozitării pe termen lung

After routine maintenance is complete, apply a neutral rust-preventive oil to the test apparatus, blades, weights, and other components to prevent rust; cover the entire unit with a dust cover and store it in a dry, well-ventilated warehouse.

5.3 Cerințe privind intervalul de calibrare

Pentru a asigura acuratețea măsurătorilor și fiabilitatea operațională a echipamentului, se recomandă trimiterea periodică a acestuia la un laborator de metrologie terț calificat pentru calibrare. Intervalul de calibrare recomandat este de 12 luni. Utilizatorii pot determina intervalul specific de calibrare în funcție de frecvența de utilizare, condițiile de mediu și cerințele sistemului de calitate.

6. Depanarea și rezolvarea problemelor comune

Simptome de eroare Cauze posibile Procedură
The blade is not perpendicular to the sample The test setup is not level. Please adjust the base of the test apparatus so that it is level.