Contor digital de energie cu 3 faze

Manual de utilizare

Aplicabil LISUN models: LS8930

 

1. Noțiuni de bază despre echipament

1.1 Specificații tehnice cheie

parametrii notițe
Intervalul de tensiune 1000 V (Settings: 1000 V, 600 V, 300 V, 150 V, 60 V, 30 V)
Gama de curent 50A (Settings: 50A, 20A, 10A, 5A, 2A, 1A)
Impedanță de intrare Voltage input impedance is approximately 2 MΩ; low-range current input impedance is approximately 4 mΩ.
Measurable Voltage Parameters Urms: True RMS; Umn: Rectified average calibrated to RMS; Udc: Simple average; Urmn: Rectified average; Uac: AC component
Measurable Current Parameters Irms: True RMS; Imn: Rectified average value calibrated to RMS; Idc: Simple average; Irmn: Rectified average; Iac: AC component
Other parameters that can be measured Cumulative energy, cumulative ampere-hours, energy time, harmonics, active power, apparent power, reactive power, power factor, phase difference, frequency, U+pk, U-pk, I+pk, I-pk, P+pk, P-pk, peak factor, WP, WP+, WP-, q, q+, q- …

1.2 Verificarea configurației echipamentului

1.2.1 Despachetarea

When unpacking the device, handle it gently to avoid scratching the casing with sharp tools. After unpacking, first inspect the exterior of the device to ensure there is no visible deformation or dents, and that the buttons are not damaged or loose. If you notice any abnormalities with the device or its accessories, do not turn it on. Please contact us immediately.

1.2.2 Unitatea principală și accesoriile

Număr notițe
LS8930 Host
Cablu de alimentare
Cablu USB
RS-485 communication cable, spare
RS-232 communication cable, spare
3 Phases Digital Power Meter-Figure1
Figura 1

1.2.3 Descărcarea documentelor electronice

We have already sent download links for the electronic user manual, software, warranty card, and other materials via email at the time of shipment. Please download them as soon as possible. If you missed the email, please feel free to contact us to request the download links again.

2. Măsuri de siguranță

2.1 Siguranța electrică

Before connecting the power cord, verify that the supply voltage is between AC 100 V and 240 V, with a frequency of 50 Hz or 60 Hz. You must use a three-prong outlet with a protective ground, and the protective ground resistance must be ≤4 Ω. Never cut or disconnect the protective ground wire, as doing so poses a risk of electric shock.

2.2 Siguranța mediului

Do not use in environments with flammable or explosive substances, strong electromagnetic interference, high temperatures and humidity, or excessive dust levels. Operating environment: Temperature 5°C to 35°C, relative humidity ≤ 65% RH.

2.3 Siguranța operațională

Înainte de cablare, the power to both the meter and the device under test must be turned off; wiring while the circuit is energized is strictly prohibited. Connect voltage wires in parallel and current wires in series; under no circumstances should a current circuit be connected to a voltage terminal or a voltage circuit to a current terminal. After wiring, tighten the terminals and ensure that no bare wires are exposed to prevent short circuits or electric shock.

Do not apply voltages or currents that exceed the rated range.

3. Equipment Overview and Usage

3.1 Introducere în panoul din spate și cablare

3.1.1 Introducere în panoul din spate

Număr notițe
Mufă pentru cablu de alimentare
Input Unit 1 Current Input Terminal
Input Unit 1 Voltage Input Terminal
Input Unit 2 Current Input Terminal
Voltage Input Terminal of Input Unit 2
Input Unit 3 Current Input Terminal
Voltage Input Terminal of Input Unit 3
RS232/RS485 interface, spare
Port USB
Ethernet port, spare
3 Phases Digital Power Meter-Figure2
Figura 2

3.1.2 Input Module Wiring Diagram

Single-phase, two-wire (1P2W), connected to input unit 1:

3 Phases Digital Power Meter-Figure3
Figura 3

Three-phase, four-wire (3P4W), connected to input terminals 1, 2, and 3:

3 Phases Digital Power Meter-Figure4
Figura 4

Three Voltages, Three Currents (3V3A):

3 Phases Digital Power Meter-Figure5
Figura 5

Three-phase, three-wire (3P3W), connected to input terminals 1 and 3:

3 Phases Digital Power Meter-Figure6
Figura 6

Single-phase, three-wire (1P3W), connected to input terminals 1 and 3:

3 Phases Digital Power Meter-Figure7
Figura 7

3.2 Introducere în panoul frontal

Număr notițe
The wiring configuration display shows the default three-phase, four-wire system. This must match the actual wiring configuration on the back panel. Press the ② wiring button, then use ③ to select the correct wiring configuration, and finally press the ④ confirm button to confirm.
The three windows on the left display three-phase voltage by default; you can switch to display other parameters using the “Unit” and “Function” buttons at position ⑥.
By default, the four windows on the right display the three-phase current and total power. You can switch to display other parameters using the unit and function keys located to the right of each window, as shown at ⑧.
3 Phases Digital Power Meter-Figure8
Figura 8

4. Utilizarea software-ului

4.1 Pregătiri

Connect the device to the computer using a USB cable. Open Device Manager on your computer; if you see the following ports listed, communication is working properly.

3 Phases Digital Power Meter-Figure9
Figura 9

4.2 Prezentare generală a software-ului

4.2.1 Software Startup Screen

Număr notițe
Select the correct port number based on the ports displayed in Device Manager.
Tap to start recording data, then tap again to stop recording.
You can export all data from the current record to an Excel spreadsheet.
Harmonic Features
Data Calculation Features
Ieșiți din software
3 Phases Digital Power Meter-Figure10
Figura 10

4.2.2 Harmonic Function Interface

Număr notițe
Click to Read Harmonic Data
Bara de progres
Message: Read successful
You can export the current harmonic data to an Excel spreadsheet.
Reveniți la ecranul de pornire
3 Phases Digital Power Meter-Figure11
Figura 11

4.2.3 Data Calculation Interface

Număr notițe
Select “Clear,” press the Settings button, and past data will be automatically cleared.
Butonul Setări
After clearing is successful, the mode automatically switches to Start; tap the Settings button again to automatically switch to Stop.
The device begins recording data
To stop, tap “Settings” again. The device will record and calculate the relevant data for this period.
3 Phases Digital Power Meter-Figure12
Figura 12
Număr notițe
Click to retrieve the relevant data into the software
Retrieval successful
You can export the current data to an Excel spreadsheet.
Reveniți la ecranul de pornire
3 Phases Digital Power Meter-Figure13
Figura 13

5. Întreținerea și îngrijirea zilnică a echipamentelor

5.1 Întreținerea de rutină

5.1.1 Curățare zilnică

Be sure to turn off the power and unplug the power cord before cleaning. We recommend wiping the unit with a soft, antistatic cotton cloth and a mild detergent; do not spray liquids directly onto the unit. Do not use chemical solvents such as benzene, toluene, or acetone.

5.1.2 Inspecții de rutină

Before each use, check that the power cord and terminal connections are secure and free of burn marks; ensure that the ventilation openings are not blocked; and after turning on the unit, verify that the indicator lights and display are functioning properly.

5.2 Specificații de stocare

For long-term storage, keep the product in a dry, dust-free, well-ventilated, and non-corrosive environment at a temperature of 5°C to 35°C and a relative humidity of ≤65% RH.

Opriți alimentarea și deconectați toate cablurile; curățați echipamentul și înfășurați-l în folie antipraf; porniți alimentarea timp de 10 de minute în fiecare lună pentru a preveni deteriorarea componentelor din cauza umezelii.

6. Depanarea și rezolvarea problemelor comune

Simptome de eroare Cauze posibile Procedură
Large Deviation in Measurement Values Wiring error; the actual wiring configuration does not match the wiring configuration selected on the device’s front panel Be sure to refer to Section 3.1.2 for the wiring scheme; be sure to refer to Section 3.2 when selecting the front panel wiring scheme.
Butonul nu funcționează Tastatura este blocată Press SHIFT+HARMO (i.e., KEY PROT) to unlock the keyboard

7. Lista de impachetat

Nu. Nume Cantitate
LS8930 Host unitate de 1
Cablu de alimentare buc 1
Cablu USB buc 1
Cablu de comunicare RS-485 buc 1
Cablu de comunicare RS-232 buc 1

Appendix A: Indicator Lights and Label Descriptions

Unit Indicator Light Descriptions

The m (mill), k (kilo), and M (giga) indicators are order-of-magnitude indicators: 1M = 1000k, 1k = 1000, 1m = 0.001

indicator luminos parametrii Unitate
V Voltaj volt
A Curent Amper
W Putere activă Țiglă
VA Putere aparentă volt-amper
Fiecare Puterea reactivă Obosit
Hz Frecvență Hertz
PF Factor de putere  
TIMP Timp Hour: Minute: Second
pk Can be combined to form a peak Determined by the group
h/h+ Can be combined to form kilowatt-hours Timp
% Relative Harmonic Content  
o Diferența de fază Grad
1 Unitatea de 1  
2 Unitatea de 2  
3 Unitatea de 3  
Σ Unitate integrată  

Indicator luminos de stare

indicator luminos notițe notițe
UPDATE Operating Status Indicator Light A steady light indicates normal operation; a flashing light indicates that data is being updated.
PĂSTREAZĂ Storage Operation Indicator Light A steady light indicates that the device is saving data; a flashing light indicates that data has been saved once.
SCALARE Magnification On Indicator Light When the light is on, it indicates that the magnification is enabled; use the magnification data for calculations.
RMS Indicator luminos de mod A lit indicator indicates that the calculated data is the effective value of the signal.
AC Indicator luminos de mod When the light is on, it indicates that the calculated data represents the AC component of the signal.
DC Indicator luminos de mod A lit indicator shows that the calculated data is the simple average of the signals.
V-MEAN Indicator luminos de mod The voltage calculation data indicated by the lit light represents the rectified average value calibrated to the effective value.
KEYPROT Button Protection A lit indicator means the button is locked and will not respond.
AVG Average Data Indicator A lit indicator means that the calculated value is the average of multiple data points.
TENSIUNE Voltage Range Indicator Light  
ACTUAL Current Range Indicator Light  
AUTO Voltage or Current Auto-Range Indicator Light When the light is on, it indicates that the voltage or current is within the range and is being automatically detected.
3P4W Wire-type indicator light The indicator light shows that the meter is a three-phase, four-wire system.
3V3A Wire-type indicator light The “Light On” Indicator Instrument Uses the 3-Voltage, 3-Current Meter Method
3P3W Wire-type indicator light The indicator light shows that the meter is a three-phase, three-wire system.
1P3W Wire-type indicator light The indicator light shows that the meter is single-phase, 3-wire.

Button Indicator Light

indicator luminos notițe notițe
MAXHOLD Indicator de valoare maximă When the light is on, it indicates that some of the calculated data values are at their maximum.
HOLD Păstrarea datelor A lit light indicates that the calculated data is held, meaning the data is no longer being updated.
START Points Accumulation Start Indicator Light When the light comes on, it indicates that the accumulation of points has begun.
STOP Point Accumulation Stop Indicator Light When the light comes on, it indicates that integration has stopped.
L-FILTER Line Filter Indicator Light A lit light indicates that the filter is on.
F-FILTER Frequency Filter Indicator Light A lit light indicates that the filter is on.
HARMO Harmonic Display On Indicator Light The light is on, indicating that the harmonic display is on.
SHIFT Second Function Indicator Light The light is on, indicating that the second function is active.