Pornirea lămpii, timpul de funcționare și sistemul de testare a pâlpâirii
Manual de utilizare
Aplicabil LISUN modele: LSRF-3
1. Noțiuni de bază despre echipament
1.1 Aplicații ale echipamentelor
1.1.1 Testarea pâlpâirii
Requires use with LISUN’s Integrating Sphere/Goniophotomoeter darkroom, DC power supply, and Standard Light Source; Existing customers can also use it directly with the LPCE-2 or LPCE-3 Integrating Sphere systems or the LSG-6000/LSG-1890B/LSG-1800A Goniophotomoeter systems.
Automatically detects parameters such as flicker frequency, flicker percentage, flicker index, modulation depth, SVM, Mp, Pst, and others; automatically classifies stroboscopic light as “no risk,” “low risk,” or “high risk” in accordance with IEEE 1789.
1.1.2 Starting the Rise Time Test
LSRF-3 trebuie folosit împreună cu LISUN’s LSP series of precision AC variable-frequency voltage stabilizers with trigger functionality, or the LSP-Pst series AC variable-frequency power supply (IEC-Pst special edition), which fully complies with the lamp startup test method specified in Section 11.4 and the lamp ramp-up test method specified in Section 11.5 of the U. S. Energy Star V2.1 guidelines, as well as the test requirements in Table 13 of SASO 2902.
Can automatically detect: startup time, rise time.
1.2 Verificarea configurației echipamentului
1.2.1 Despachetarea
When unpacking the device, handle it gently to avoid scratching the casing with sharp tools. After unboxing, first inspect the exterior of the device to ensure there is no visible deformation or dents, and that the buttons are not damaged or loose. If you notice any abnormalities with the device or its accessories, do not turn it on. Please contact us immediately.
1.2.2 Unitatea principală și accesoriile
Vezi Figura 1-1.
| Număr | notițe |
| ① | LSRF-3 Unitatea principala |
| ② | Adaptor |
| ③ | Sondă de luminanță |
| ④ | Trigger cable, used to connect to LISUN AC power supplies with a trigger function |
| ⑤ | Probe connection cable, used to connect the main unit to the photometer probe |
| ⑥ | Temperature and humidity probe, used to measure temperature and humidity in a laboratory |
| ⑦ | USB cable, used to connect the console to a computer |

Figura 1-1
1.2.3 Descărcarea documentelor electronice
We have already sent you the download links for the electronic user manual, software, warranty card, and other materials via email at the time of shipment. Please download them as soon as possible. If you have not received the email, please contact us to request the download links again.
2. Precauții pentru utilizare
2.1 Siguranța electrică
Before connecting the power cord, verify that the supply voltage is AC 220 V ±10% and the frequency is 50 Hz or 60 Hz. You must use a three-prong outlet with a protective ground, with a protective ground resistance of ≤4 Ω. Never cut or disconnect the protective ground wire, as doing so poses a risk of electric shock.
2.2 Siguranța mediului
Do not use in environments with flammable or explosive substances, strong electromagnetic interference, high temperatures and humidity, or excessive dust levels. Operating environment: Temperature 18°C to 30°C, humidity 35% RH to 75% RH.
2.3 Proceduri de operare
Low-light tests must completely block out stray light; Integrating Sphere must be sealed and free of light leaks. For darkroom tests, ensure that the probe is directly facing the light source and that there are no interfering reflective light sources in the surrounding area. Test samples must be placed in a stationary position to avoid fluctuations in light output caused by vibration. As the unit contains low-voltage precision components, do not disassemble the main unit’s casing without authorization.
3. Prezentare generală a echipamentului și cablare
3.1 Introducere în panoul frontal al echipamentului
Vezi Figura 3-1.
| Număr | notițe |
| ① | Întrerupător |
| ② | Equipment Status Indicator Lights |
| ③ | Touchscreen. Note: All device operations are performed by computer software; no interaction with the touchscreen is required. |

Figura 3-1
3.2 Wiring the LSRF-3 Host to the LSP Series Power Supply
Vezi Figura 3-2.
| Număr | notițe |
| ① | Connect the main unit to the computer using a USB cable to enable software communication and power the device via the USB cable. Note: You may also connect the included power adapter at the same time to ensure the device continues to function properly even if the USB power supply is unstable. |
| ② | Connect the temperature and humidity probe; place the sensor in an unobstructed area of the laboratory to measure the ambient temperature and humidity. |
| ③ | Connect the photometric probe; connect the probe end to Integrating Sphere or a darkroom. |
| ④ | Connect the trigger wire; connect the other end to the LSP series power supply. |
| ⑤ | Connect the trigger wire; connect the other end to the LSRF-3. |
| ⑥ | Connect the RS-232 communication cable; connect the other end to a computer or to a computer via an RS-232-to-USB adapter cable. |
| ⑦ | Plug in Input power supply cord to connect the LSP to Input power supply |
| ⑧ | Output power connector; connect to the sample to supply power to it |

Figura 3-2
Note: If an LSP series power supply is not configured, there is no need to connect the trigger line; however, rise time tests cannot be performed, nor can the power supply output be automatically controlled via software.
3.3 Connecting the Photometric Probe
3.3.1 Integrating Sphere Scenario
When used with Integrating Sphere, see Figure 3-3.

Figura 3-3
3.3.2 Darkroom Scenes
When using the device in a darkroom, use the standard probe holder to fix the probe’s orientation so that it is directly aligned with the center of the light source being measured, as shown in the schematic in Figure 3-4.

Figura 3-4
4. Software Overview
4.1 Instalarea software-ului
Software Runtime Environment: Windows 7 or later.
Consultați Figura 4-1 pentru pachetul de instalare a software-ului.
| Număr | notițe |
| ① | LSRF-3 Software Installer—please double-click to install |
| ② | LSRF-3 Device Driver—Double-click to install |
| ③ | If the LSP series power supply is connected to a computer via an RS232-to-USB adapter cable, you will also need to install the USB adapter driver. |

Figura 4-1
4.2 Introduction to Stroboscopic Testing Software
Vezi Figura 4-2.
| Număr | notițe |
| ① | Files: New, Open, Save, Save As, Export/Import CSV/TXT files, Print, Print Preview, Exit |
| ② | Settings: System Configuration, Product Information, Photometric Calibration |
| ③ | Testing: Single Test, Continuous Test, Batch Test, Stop Test |
| ④ | Help: Software Version Number, Switch Between Chinese and English |
| ⑤ | New: Create a new blank test file in .lsfx format |
| ⑥ | Open: Open a saved test file in .lsfx format |
| ⑦ | Save: Save the current file settings |
| ⑧ | Imprimați raportul de testare |
| ⑨ | Informaţii despre produs |
| ⑩ | Configurația de sistem |
| ⑪ | Photometric Calibration |
| 10 | Test unic |
| 10 | Testarea continuă |
| 10 | Test pe lot |
| ⑮ | Opriți testarea |
| ⑯ | Număr versiune software |
| ⑰ | X-axis: Time; Y-axis: Luminous flux |
| 11 | Horizontal axis time range. Enter manually and confirm; Auto-select: The software automatically selects a range; Show All |
| ⑲ | Click the “Standards” tab to view the test spectra for the corresponding standards. |
| ⑳ | All test information and test result data are displayed |

Figura 4-2
4.3 Introduction to the Start-Up and Ramp-Up Test Software
Vezi Figura 4-3.
| Număr | notițe |
| ① | File: New, Open, Save, Save As, Export to CSV/TXT, Print, Print Preview, Exit |
| ② | Settings: System Configuration, Test Settings, Product Information, Criteria Settings, Photometric Calibration |
| ③ | Test: Test, Stop Test |
| ④ | Help: Software Version Number, Switch Between Chinese and English |
| ⑤ | New: Create a new blank test file in .lsrx format |
| ⑥ | Open: Open a saved test file in .lsrx format |
| ⑦ | Save: Save the current file settings |
| ⑧ | Imprimați raportul de testare |
| ⑨ | Informaţii despre produs |
| ⑩ | Configurația de sistem |
| ⑪ | Configurarea testului |
| 10 | Photometric Calibration |
| 10 | Testare |
| 10 | Opriți testarea |
| ⑮ | Număr versiune software |
| ⑯ | Time-course plot; horizontal axis: time; vertical axis: luminous flux |
| ⑰ | Horizontal axis time range: Manually enter and confirm, auto-select, or display all; Get start time: Click to automatically retrieve the start time; Custom start time: Not recommended |
| 11 | Rise-time graph; horizontal axis: time; vertical axis: luminous flux |
| ⑲ | Horizontal axis time range: Manual entry followed by confirmation, automatic selection, or “Show All”; Calculate rise time: Click to automatically calculate the start time; Custom rise time: Not recommended |
| ⑳ | Test information shows |

Figura 4-3
5. Steps for Stroboscopic Testing
5.1 Setări de configurare a sistemului
5.1.1 Communications
Vezi Figura 5-1.
| Număr | notițe |
| ① | Connection Type: Select USB |
| ② | Device Serial Number: Automatically retrieved by the software |
| ③ | Photometric Type: Select “Luminous Flux” when using with Integrating Sphere; select “Illuminance” or “Luminous Intensity” when using in a darkroom. |
| ④ | Test Distance: Enter the correct test distance when using the device in a darkroom; this parameter is not applicable when using the device with Integrating Sphere. |
| ⑤ | AC Power Supply Model: If you have purchased an LISUN LSP Series or LSP-PST Series AC power supply, select the correct model |
| ⑥ | AC Power Communication Port: Select “Auto” to automatically detect the port. |
| ⑦ | Once you’ve finished configuring everything, click “OK” to save and exit. |

Figura 5-1
Afișajul 5.1.2
Vezi Figura 5-2.
| Număr | notițe |
| ① | Waveform display range after data is loaded: Select either “Auto” or “All” |
| ② | Number of Automatic Display Cycles: When “Automatic” is selected, the number of stroboscopic cycles the software automatically captures |
| ③ | The starting point on the vertical axis is always 0: You can check or uncheck this box as needed. |
| ④ | CIE SVM Graph Display: “Weighted Relative Spectral Energy” is selected by default |
| ⑤ | ASSIST Graph Display: “Spectral Perceptible Flicker” is selected by default |
| ⑥ | Minimum frequency: 5 Hz (default) |
| ⑦ | Once you’ve finished configuring everything, click “OK” to save and exit. |

Figura 5-2
5.1.3 Imprimare
Vezi Figura 5-3.
| Număr | notițe |
| ① | Language Selection: Chinese or English |
| ② | Report Content Selection |
| ③ | Waveform Capture Range Selection: Auto, All, Match Current Display |
| ④ | Setări antet |
| ⑤ | Should a divider line appear below the header? |
| ⑥ | Once you’ve finished configuring everything, click “OK” to save and exit. |

Figura 5-3.
Stocare 5.1.4
Vezi Figura 5-4.
| Număr | notițe |
| ① | Whether to enable auto-save for test files; we recommend checking this box |
| ② | File Save Location |
| ③ | Convenții de denumire a fișierelor |
| ④ | Enable automatic saving of test parameters to a list file. We recommend checking this box. |
| ⑤ | File Save Location |
| ⑥ | Enter the file name. The file format must be .csv. |
| ⑦ | Once you’ve finished configuring everything, click “OK” to save and exit. |

Figura 5-4
5.2 Photometric Calibration
5.2.1 Procedura de calibrare
As shown in Figure 5-5, the calibration procedure is as follows.
| Număr | notițe |
| ① | Integrating Sphere: Ensure there is no light inside Integrating Sphere, and close Integrating Sphere; Darkroom: Ensure there is no light inside the darkroom, and close it. Then click “Zero Calibration,” wait for the software to indicate that zero calibration was successful, and confirm to exit. |
| ② | Integrating Sphere: Select “luminous flux”; Darkroom: Select “illuminance” |
| ③ | Integrating Sphere: Install the Standard Light Source exactly at the center of the Integrating Sphere and turn it on; close the Integrating Sphere and enter the luminous flux of the Standard Light Source; Darkroom: Mount the Standard Light Source on the Goniophotomoeter and turn it on. Rotate the Goniophotomoeter so that the Standard Light Source is directly facing the LSRF-3 photometric probe. Close the darkroom and enter the illuminance value of the Standard Light Source. |
| ④ | We recommend waiting 10 to 15 minutes for Standard Light Source to stabilize, then clicking “Calibrate.” |

Figura 5-5
5.2.2 Frecvența de calibrare
Failure to calibrate does not affect the stroboscopic test results, but photometric data such as luminous flux may be inaccurate. If you wish to obtain accurate photometric data (such as luminous flux) or if you find that the photometric data is inaccurate, you can perform a calibration.
5.3 Pregătiri
5.3.1 Installing Lights
| Scenă | notițe |
| Integrarea sferei | Install the luminaire under test at the center of Integrating Sphere, and close Integrating Sphere |
| darkroom | Mount the luminaire under test on the Goniophotomoeter, rotate the Goniophotomoeter so that its light-emitting surface faces the probe, and close the darkroom. |
5.3.2 Reference for Sampling Parameters Under Different Standards
| Standarde aplicabile | Rata de eșantionare | Timp de prelevare |
| CA CEC | ≥20 kS/s | ≥1 s |
| CIE SVM | ||
| ASISTA | ≥10 kS/s | ≥2 s (≥10 s for unstable light sources) |
| IEC-Pst | ≥180 secunde | |
| Energy Star V2.1 | ≥10 kS/s (Pst/Mp), ≥20 kS/s (all others) | ≥180 s (Pst/Mp) |
| IEEE Std 1789 | ≥2 kS/s | ≥ 8 cycles of the signal under test |
| BASIC | ≥10 kS/s | ≥5 secunde |
When performing the Pst LM (I) test in accordance with IEC TR 61547-1 (IEC-Pst), you must select one of the five built-in sets of modulation frequencies and fluctuation percentages provided by the software; custom settings are not permitted. Modulation is not required for other tests or other standards.
5.4 Începeți testarea
5.4.1 Single-Test/Continuous-Test Interface
Vezi figurile 5–6.
| Număr | notițe |
| ① | If you select “Unmodulated,” enter the Input power supply voltage and frequency for the sample luminaire. |
| ② | If you select “Pst Power Modulation Mode,” choose the Input power supply voltage and frequency for the test luminaire, and then select the modulation ratio . Note: This feature requires a LISUN LSP-PST series AC power supply. |
| ③ | Click “Output” to control the power supply output. Check the boxes for “Automatically output at the start of the test” and “Stop output at the end of the test” to allow the software to automatically control the power supply output . Note: This feature requires a LISUN LSP/LSP-PST series AC power supply. |
| ④ | Sampling Rate Settings |
| ⑤ | Sampling Time Settings |
| ⑥ | To set the measurement range, simply select “Auto Range.” |
| ⑦ | Warm-up Time Setting: This setting is used to test the flicker data of the lighting fixture in a stable state. After clicking “Start,” the software begins a countdown, and the test will run automatically once the countdown ends. |
| ⑧ | Interval setting—applies only to continuous testing, i.e., the interval between two consecutive tests |
| ⑨ | Use trigger-based testing; this is a backup feature—please do not check this box. |
| ⑩ | Once everything is set up, click OK to start the test. |

Figurile 5–6
Note: During continuous testing, all reports will be automatically saved according to the settings in 5.1.4.
5.4.2 Batch Testing Interface
Vezi figurile 5–7.
| Număr | notițe |
| ① | Select a sample light fixture and enter the Input power supply voltage |
| ② | Check the “Stop Output at End of Test” box, and the software will automatically turn off the power output when the test ends. |
| ③ | Whether modulation is required at 50 Hz and the modulation ratio; these options can be selected in bulk |
| ④ | Whether modulation is required at 60 Hz and the modulation ratio; these options can be selected in bulk. |
| ⑤ | Test Intervals for Different Modulations |
For other settings, see Section 5.3.2.

Figura 5-7
Note: For batch testing, all reports will be automatically saved according to the settings in 5.1.4.
5.5 Viewing Test Reports
Refer to Section 4.2 to open the “Product Information” screen and enter the information, as shown in Figure 5-8.
| Număr | notițe |
| ① | Introduceți rapoarte, produse și informații despre teste |
| ② | The software automatically measures the temperature and humidity of the test environment. |
| ③ | Reports, Auto-Increment Feature for Product Numbers |
| ④ | Once you have filled in and configured all the settings, click “OK” to save and exit. |

Figura 5-8
Then, following the instructions in Section 4.2, print the test report or export the test data.
6. Steps for Performing the Start-up Rise Time Test
6.1 Setări de configurare a sistemului
6.1.1 Configurare
Vezi Figura 6-1.
| Număr | notițe |
| ① | Selectați modelul LSRF-3; the software will automatically retrieve the serial number. |
| ② | Select the AC power supply model and choose “Auto Detect” for the port; the software will automatically detect the communication port. |
| ③ | The options “Automatically Turn On Power” and “Automatically Turn Off Power” are checked by default. The software must automatically control the power supply to run the ramp-up test; otherwise, the test cannot be performed. |
| ④ | For the sync signal, “Rising Edge Trigger” is selected by default; “Falling Edge Trigger” is a backup function that is currently unavailable—please do not select it. |
| ⑤ | Custom Observation Points for Rise Characteristics: You can enter a specific time to view the luminous flux value at that moment. |
| ⑥ | To display the voltage waveform in the characteristic waveform display window when it is launched, you can check or uncheck this box as needed. |
| ⑦ | Once you’ve finished configuring everything, click “OK” to save and exit. |

Figura 6-1
6.1.2 Imprimare
See 5.1.3; the functionality is the same.
Stocare 6.1.3
See Section 5.1.4; the functionality is the same.
6.2 Photometric Calibration
6.2.1 Procedura de calibrare
See Figure 6-2; the calibration procedure is as follows.
| Număr | notițe |
| ① | Integrating Sphere: Ensure there is no light inside Integrating Sphere, and close Integrating Sphere; Darkroom: Ensure there is no light inside the darkroom, and close it. Then click “Zero Calibration,” wait for the software to indicate that zero calibration was successful, and confirm to exit. |
| ② | According to the standard, luminous flux must be used as a reference for all rise time tests; therefore, whether using Integrating Sphere or a darkroom, calibration must be performed using luminous flux. |
| ③ | Integrating Sphere: Install the Standard Light Source exactly at the center of the Integrating Sphere and turn it on; close the Integrating Sphere and enter the luminous flux of the Standard Light Source; Dark Chamber: Mount the luminaire under test on the Goniophotomoeter and turn it on; rotate the Goniophotomoeter so that its light-emitting surface faces the LSRF-3 photometric probe; close the dark chamber; enter the luminous flux of the luminaire under test. |
| ④ | We recommend waiting 10 to 15 minutes until the light source stabilizes, then clicking “Calibrate.” |

Figura 6-2
6.2.2 Frecvența de calibrare
Failure to calibrate does not affect the results of the startup rise time test, but the photometric data will be inaccurate. If you want to obtain accurate photometric data, such as luminous flux, or if you find that the photometric data is inaccurate, you can run a calibration.
6.3 Pregătiri
6.3.1 Installing Lights
| Scenă | notițe |
| Integrarea sferei | Install the luminaire under test at the center of Integrating Sphere, and close Integrating Sphere |
| darkroom | Mount the luminaire under test on the Goniophotomoeter, rotate the Goniophotomoeter so that its light-emitting surface faces the probe, and close the darkroom. |
6.3.2 Stabilirea criteriilor
Refer to Section 4.3 to access the criteria settings interface, as shown in Figure 6-3.
| Număr | notițe |
| ① | Define the startup time as a percentage; for example, based on Energy Star guidelines, you can set it to 98%. |
| ② | Defining the Start Time Based on Peak Values |
| ③ | Customizable; please contact us if you have any other customization requests. |
| ④ | Define the rise time as a percentage; for example, based on Energy Star guidelines, it can be set to 80%. |
| ⑤ | Definition of lamp stability: For example, within the specified stabilization period of 15 minutes, the fluctuation in light output does not exceed 1%. |
| ⑥ | După ce ați terminat de configurat setările, faceți clic pe OK pentru a salva și a ieși. |

Figura 6-3
6.4 Începeți testarea
Click “Test” to open the test settings screen, as shown in Figure 6-4.
| Număr | notițe |
| ① | Set the power output parameters: voltage, frequency, and initial phase angle of the supply voltage |
| ② | Click to automatically select the correct range |
| ③ | Start Time Test Parameter Settings: Sampling Duration, Sampling Rate |
| ④ | If you need to test the rise time at the same time, check the “Test Rise Time” box, and then set the test parameters: rise time test duration and rise time test interval. |
| ⑤ | Once you’ve finished configuring the settings, click “OK” to start the test automatically. |

Figura 6-4
6.5 Viewing Test Reports
Refer to Section 4.3 and open the “Product Information” screen, as shown in Figure 6-5.
| Număr | notițe |
| ① | Introduceți rapoarte, produse și informații despre teste |
| ② | Reports, Auto-Increment Function for Product IDs |
| ③ | Once you have filled in and configured all the settings, click “OK” to save and exit. |

Figura 6-5
Then, following the instructions in Section 4.3, print the test report or export the test data.
7. Întreținerea și îngrijirea zilnică a echipamentelor
7.1 Întreținerea de rutină
7.1.1 Curățare zilnică
Be sure to turn off the power and unplug the power cord before cleaning. We recommend wiping the device with a soft, antistatic cotton cloth and a mild detergent; do not spray liquids directly onto the device. Do not use chemical solvents such as benzene, toluene, or acetone.
7.1.2 Inspecții de rutină
Before each use, check the power cord and terminal connections to ensure there are no loose connections or signs of burn damage; after turning on the unit, check that the indicator lights and display are functioning properly.
7.2 Maintenance Guidelines Specific to Photometric Probes
The probe is a precision optical device; do not bump, drop, or expose it to direct strong light. Never touch the probe’s light-sensitive surface with your bare hands; gently wipe away dust with lint-free lens paper. If the probe will not be used for an extended period, secure the probe cap, place it in a moisture-proof storage case, and store it in a dry, cool place.
7.3 Specificații de stocare
7.3.1 Cerințe privind mediul de stocare
Temperature: 5°C to 35°C; humidity: ≤75% RH; dry, dust-free environment free of corrosive gases.
7.3.2 Operațiuni de depozitare pe termen lung
Opriți alimentarea și deconectați toate cablurile; curățați echipamentul și înfășurați-l în folie antipraf; porniți alimentarea timp de 10 de minute în fiecare lună pentru a preveni deteriorarea componentelor din cauza umezelii.
7.4 Cerințe privind intervalul de calibrare
Pentru a asigura acuratețea măsurătorilor și fiabilitatea operațională a echipamentului, se recomandă trimiterea periodică a acestuia la un laborator de metrologie terț calificat pentru calibrare. Intervalul de calibrare recomandat este de 12 luni. Utilizatorii pot determina intervalul specific de calibrare în funcție de frecvența de utilizare, condițiile de mediu și cerințele sistemului de calitate.
8. Depanarea și rezolvarea problemelor comune
8.1 LSRF-3 Stroboscopic Start-Up Rise Tester
| Simptome de eroare | Cauze posibile | Procedură |
| The device frequently restarts on its own, and the screen flickers | Insufficient or Unstable USB Power Supply | While connecting the USB cable, plug the power adapter into the device. |
| Software lags or closes unexpectedly | Insufficient computer permissions | Please log in to your computer using an administrator account and try again. |
| Inaccurate photometric test parameters | Not calibrated or calibration failed | Try again after calibrating according to Section 5.2 or 6.2. |
8.2 LSP Series Power Supplies
| Simptome de eroare | Cauze posibile | Procedură |
| Nu se poate porni | Defecțiune a comutatorului selectorului de tensiune de intrare, siguranță arsă | Verificați comutatorul selectorului de tensiune de intrare; opriți alimentarea, eliminați orice scurtcircuit sau supraîncărcare la ieșire, înlocuiți siguranța cu una de aceeași valoare și nu creșteți curentul nominal. |
| Ventilatorul nu se învârte | Acumulare de praf, defecțiune a motorului | Opriți alimentarea și curățați praful de pe ventilator; dacă problema persistă, vă rugăm să ne contactați pentru reparații sau pentru înlocuirea ventilatorului. |
| Codul de eroare 02 în colțul din dreapta jos al afișajului | In the event of an output short circuit, excessive current, or excessive power, the power supply will automatically cut off the output to protect itself. | Please turn off the LSP power supply, check the wiring, and resolve any output short circuits before restoring power; if the load current or power exceeds the power supply’s upper limit, you will need to purchase a larger AC power supply. |
| Codul de eroare 01 în colțul din dreapta jos al afișajului | S-a ars o siguranță | Please turn off the LSP power supply and unplug the power cord. Check all fuses on the rear panel of the power supply and replace them with fuses of the same rating. Do not increase the current rating. |

中文简体